Search criteria

1 vulnerability found for Zeus RS C500 by Dräger

CVE-2025-15653 (GCVE-0-2025-15653)

Vulnerability from cvelistv5 – Published: 2026-06-02 21:27 – Updated: 2026-06-03 20:01
VLAI?
Title
Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation
Summary
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
CWE
  • CWE-668 - Exposure of Resource to Wrong Sphere
Assigner
Impacted products
Vendor Product Version
Dräger Zeus IE Affected: 0 , < 1.0.5 (custom)
Create a notification for this product.
    Dräger Zeus RS C500 Affected: 0 , < 1.0.9 (custom)
Create a notification for this product.
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2025-15653",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "total"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2026-06-03T14:05:59.046177Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2026-06-03T14:06:48.680Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unknown",
          "product": "Zeus IE",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "lessThan": "1.0.5",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        },
        {
          "defaultStatus": "unknown",
          "product": "Zeus RS C500",
          "vendor": "Dr\u00e4ger",
          "versions": [
            {
              "lessThan": "1.0.9",
              "status": "affected",
              "version": "0",
              "versionType": "custom"
            }
          ]
        }
      ],
      "datePublic": "2025-12-15T00:00:00.000Z",
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003eDr\u00e4ger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dr\u00e4ger Service Connect.\u003c/p\u003e"
            }
          ],
          "value": "Dr\u00e4ger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dr\u00e4ger Service Connect."
        }
      ],
      "metrics": [
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "NONE",
            "attackVector": "PHYSICAL",
            "baseScore": 7,
            "baseSeverity": "HIGH",
            "exploitMaturity": "NOT_DEFINED",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "HIGH",
            "vulnConfidentialityImpact": "HIGH",
            "vulnIntegrityImpact": "HIGH",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "PHYSICAL",
            "availabilityImpact": "HIGH",
            "baseScore": 6.8,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-668",
              "description": "CWE-668 Exposure of Resource to Wrong Sphere",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2026-06-03T20:01:38.486Z",
        "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "shortName": "VulnCheck"
      },
      "references": [
        {
          "tags": [
            "vendor-advisory"
          ],
          "url": "https://static.draeger.com/security/download/Product-Security-Advisory-25-349-Zeus.pdf"
        },
        {
          "tags": [
            "third-party-advisory"
          ],
          "url": "https://www.vulncheck.com/advisories/dr-ger-zeus-ie-anesthesia-workstation-usb-interface-privilege-escalation"
        }
      ],
      "source": {
        "discovery": "INTERNAL"
      },
      "title": "Dr\u00e4ger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation",
      "x_generator": {
        "engine": "Vulnogram 1.0.2"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
    "assignerShortName": "VulnCheck",
    "cveId": "CVE-2025-15653",
    "datePublished": "2026-06-02T21:27:37.997Z",
    "dateReserved": "2026-06-02T21:25:08.662Z",
    "dateUpdated": "2026-06-03T20:01:38.486Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.2"
}