CVE-2025-15653 (GCVE-0-2025-15653)
Vulnerability from cvelistv5 – Published: 2026-06-02 21:27 – Updated: 2026-06-03 20:01
VLAI?
Title
Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation
Summary
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
Severity ?
CWE
- CWE-668 - Exposure of Resource to Wrong Sphere
Assigner
References
| URL | Tags | |||||||
|---|---|---|---|---|---|---|---|---|
|
||||||||
Impacted products
| Vendor | Product | Version | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Dräger | Zeus IE |
Affected:
0 , < 1.0.5
(custom)
|
|||||||
|
|||||||||
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2025-15653",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-06-03T14:05:59.046177Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T14:06:48.680Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "unknown",
"product": "Zeus IE",
"vendor": "Dr\u00e4ger",
"versions": [
{
"lessThan": "1.0.5",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
},
{
"defaultStatus": "unknown",
"product": "Zeus RS C500",
"vendor": "Dr\u00e4ger",
"versions": [
{
"lessThan": "1.0.9",
"status": "affected",
"version": "0",
"versionType": "custom"
}
]
}
],
"datePublic": "2025-12-15T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\u003cp\u003eDr\u00e4ger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dr\u00e4ger Service Connect.\u003c/p\u003e"
}
],
"value": "Dr\u00e4ger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dr\u00e4ger Service Connect."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "NONE",
"attackVector": "PHYSICAL",
"baseScore": 7,
"baseSeverity": "HIGH",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "HIGH",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "PHYSICAL",
"availabilityImpact": "HIGH",
"baseScore": 6.8,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-668",
"description": "CWE-668 Exposure of Resource to Wrong Sphere",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-06-03T20:01:38.486Z",
"orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"shortName": "VulnCheck"
},
"references": [
{
"tags": [
"vendor-advisory"
],
"url": "https://static.draeger.com/security/download/Product-Security-Advisory-25-349-Zeus.pdf"
},
{
"tags": [
"third-party-advisory"
],
"url": "https://www.vulncheck.com/advisories/dr-ger-zeus-ie-anesthesia-workstation-usb-interface-privilege-escalation"
}
],
"source": {
"discovery": "INTERNAL"
},
"title": "Dr\u00e4ger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation",
"x_generator": {
"engine": "Vulnogram 1.0.2"
}
}
},
"cveMetadata": {
"assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"assignerShortName": "VulnCheck",
"cveId": "CVE-2025-15653",
"datePublished": "2026-06-02T21:27:37.997Z",
"dateReserved": "2026-06-02T21:25:08.662Z",
"dateUpdated": "2026-06-03T20:01:38.486Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
Loading…
Loading…
Sightings
| Author | Source | Type | Date |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…